Herunterladen Inhalt Inhalt Diese Seite drucken

BBraun Aesculap Acculan 3Ti Gebrauchsanweisung Seite 18

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 28
Aesculap
®
Acculan 3Ti
Acculan 3Ti small drill GA671
Validated cleaning and disinfection procedure
Validated procedure
Manual cleaning and
wipe disinfection
16
Special features
When cleaning instruments with
movable hinges, ensure that these
are in an open position and, if
applicable, move the joint while
cleaning.
Drying phase: Use a lint-free cloth
or compressed air for medical
purposes
Ensure that the product is posi-
tioned in such a way that water
will not enter the product e.g.
through coupling interfaces.
(Immediately remove any fluid
that entered the product inadver-
tently.)
For cleaning the cannulation, use
special cleaning brush
(TA011944).
Clean difficult-to-access areas
with soft round plastic brushes of
fitting diameter.
Reference
Chapter Manual cleaning/disinfecting and
subsection:
Chapter Manual cleaning and wipe dis-
infecting