To deactivate the oscillating or tapping mode:
Press counterclockwise rotation button 7 for 3 s,
see Fig. 12.
The drill issues an acoustic signal.
7
8
9
6
Fig. 12
5.
Validated reprocessing pro-
cedure
5.1

General safety instructions

Note
Adhere to national statutory regulations, national and
international standards and directives, and local, clini-
cal hygiene instructions for sterile processing.
Note
For patients with Creutzfeldt-Jakob disease (CJD), sus-
pected CJD or possible variants of CJD, observe the rel-
evant national regulations concerning the reprocessing
of products.
Note
Mechanical reprocessing should be favored over man-
ual cleaning as it gives better and more reliable results.
Note
Successful processing of this medical device can only
be ensured if the processing method is first validated.
The operator/sterile processing technician is responsi-
ble for this.
The recommended chemistry was used for validation.
Note
If there is no final sterilization, then a virucidal disin-
fectant must be used.
Note
For the latest information on reprocessing and material
compatibility see also the Aesculap extranet at
www.extranet.bbraun.com
The validated steam sterilization procedure was carried
out in the Aesculap sterile container system.