Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 24

TROUBLESHOOTING

SYMPTOM
A. Motor does not rotate.
Display does not light.
B. Motor does not rotate.
Display lights.
CAUSE
A. No power.
B1. Defective remote control.
B2. MODE control not
properly set.
REMEDY
1. Check fuse and replace if
defective (Fig. 3).
2. Check that unit is plugged into
a live line.
3. Check connection of power
cord.
4. Check the line cord for continuity
and replace if defective.
5. Return for servicing.
1. Place power switch in
off position.
2. Check that remote cable
connector is inserted fully into
the receptacle (Fig. 2 and 3)
3. If motor still does not rotate,
select INTernal with the MODE
control and press the
START/STOP control (Fig. 1).
4. If the motor rotates, replace the
remote control with similar unit.
If motor does not rotate, return
drive for servicing.
1. Check that the MODE control
is set to INT for operation with
front panel control or to mA
or V for operation with remote
control (Fig. 1).
2. If motor still does not rotate,
return for servicing.
Troubleshooting continued on page 10.
9

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis