ASPINA DO M INSTALLATION – OPERATION – MAINTENANCE
Guidance and manufacturer's declaration - electromagnetic immunity
Pursuant to IEC 60601-1-2:2014 - Medical electrical equipment - Part 1-2: General requirements for
basic safety and essential performance - Collateral Standard: Electromagnetic disturbances
The equipment is intended for use in the electromagnetic environment specified below. The customer
or the user of the equipment should assure that it is used in such an environment.
IEC 60601-1-2
Immunity test
Electrostatic
±8 kV contact
discharge (ESD)
±15 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply
IEC 61000-4-4
lines
±1 kV for
input/output lines
Surge
±1 kV differential
IEC 61000-4-5
mode
±2 kV common
mode
Voltage dips,
U
=0%, 0,5 cycle
T
short interruption,
(at 0, 45,
and voltage
90, 135, 180,
variations on
225, 270 and
power
315°)
supply
input lines IEC
U
=0%, 1 cycle
T
60601-4-11
U
=70% 25/30
T
cycles (at 0°)
U
=0%, 250/300
T
cycles
NOTE: U
is the A.C. mains voltage prior to application of the test level.
T
Compliance level
test level
±8 kV contact
±15 kV air
±2 kV
100 kHz repetition
frequency
Applied to mains
connection
±1 kV L-N
±2 kV L-PE; N-PE
Applied to mains
connection
U
=>95%, 0,5 cycle
T
(at 0, 45, 90, 135,
180, 225, 270 and
315°)
U
=>95%, 1 cycle
T
U
=70% (30% dipp
T
U
),
T
25(50Hz)/30(60Hz)
cycles (at 0°)
U
=>95%,
T
250(50Hz)/300(60Hz)
cycle
Electromagnetic environment -
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
The device stops and restarts
automatically at each dip. In this
case does not occur unacceptable
pressure drop.
- 8 -
guidance
NP-DO M-34_03-2019