Errors
Error
RTC MISSING
RTC CRC
CHECK CONFIG
TOO MANY PROBES
TOO MANY
MEASUREMENTS
RTC FAILURE
SYNTAX ERROR
FORMULA TO LONG
ARGUMENT
LOGIC FUNCTION
BOUNDARY
FUNCTION

Replacement parts and accessories

Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Description
DIN rail NS 35/15 (DIN EN 60715 TH35), galvanized steel, 35 cm
(13.78 in.) length
Transformer, 90–240 V AC, 24 V DC, 0.75 A (for DIN rail NS 35/15)
Terminal block for 24 V connection, without power supply
Description
There is no communication
between RTC module and
RTC communication card.
The communication between
RTC module and RTC
communication card was
cancelled.
The sensor selection of the
RTC module was removed by
removal or selection of a new
sc1000 controller.
Too many sensors selected in
the SELECT SENSOR menu.
The probes selected in the
SELECT SENSOR have too
many measurements to be
operates by the RTC
communication card.
A general reading/writing error
of on CF card, which was most
likely caused by a brief
interruption to the power
supply.
Error in PROGNOSYS *.bin-
file.
W A R N I N G
Solution
Supply RTC module with voltage.
Examine the connection cable.
Set the power of the sc1000 and the RTC module to
off. Wait until the system is completely voltage free. Set
the power of the sc1000 controller and the RTC module
to on.
Make sure that the +/- connections of the connector
cable between RTC and RTC communication card in
the sc1000 are installed correctly. Make the necessary
changes.
Go to: MAIN MENU>RTC
MODULES/PROGNOSYS>RTC MODULES>RTC>
CONFIGURE>SELECT SENSOR, select the correct
sensor for the RTC again and confirm.
Go to the SELECT SENSOR menu. Select no more
than 15 probes.
Go to the SELECT SENSOR menu. Select the number
of probes that have no more than 15 measurement
values.
Acknowledge the error. If the message frequently
shows, remove the cause of power disruptions. If
necessary, contact technical support.
Use an updated version of PROGNOSYS that does not
show this error. Contact technical support.
Quantity
1
1
1
Item no.
LZH165
LZH166
LZH167