Herunterladen Inhalt Inhalt Diese Seite drucken

Mettler Toledo M300 Schnellkonfigurationsanleitung Seite 10

Vorschau ausblenden Andere Handbücher für M300:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 17
Setpoints
7.0
pH
25.0
°C
a Set Point Yes
SP1 Type= Off
7.0
pH
25.0
°C
SP1 BetweenH= 0.000
SP1 BetweenL= 0.000
7.0
pH
25.0
°C
SP1 use Relay #1
10
After configuring the Analog Output a setpoint can be configured for that output. If No
is selected and [ENTER] is pressed then the quick setup is done and the menus are
exited without setting up any setpoint.
Selecting Yes and pressing [ENTER] means a setpoint can be configured for channel a.
Following setpoint Types can be selected:
– High (High value has to be set)
– Low (Low value has to be set)
– Between (High and Low value has to be set)
– Outside (High and Low value has to be set)
For Mettler-Toledo Thornton Conductivity only:
– USP (% safety margin below U.S. Pharmacopoeia limits)
– EP PW (% safety margin below European Pharmacopeia limits for Purified Water)
– EPWFI (% safety margin below European Pharmacopeia limits for Water for Injection)
After setting the setpoint value(s) a Relay (none, 1, 2, 3, 4) can be configured for
that setpoint. The Relay delay is set to 10 seconds and the Hysteresis is set to 5%. If a
relay is configured as CLEAN relay, it will not be selectable in this menu.
For multi-channel devices: Also setpoints 5 and 6 can be configured and the user can
configure a setpoint to measurement c by going back thourgh previous menus and se-
lecting c.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis