Pressure error
Waste full
File error
Dynamic memory
error
56V power error
Problembehebung
The pressure inside the
pressure chamber does not
reach the expected value
within the given time
The waste container is full.
Something is wrong with the
analyzer's file system.
Something is wrong with the
analyzer's memory.
Something is wrong with the
power board.
11-11
If the test result is normal, the error will
be removed;
If the problem remains, contact Mindray
customer service department or your
local distributor.
Check whether the external tubing is
pressed;
If not, enter "Service→ Self-test →
Tubing"screen to check the pressure
as instructed by Chapter 10.5..;
If the test result is normal, the error will
be removed.
If the problem remains, contact Mindray
customer service department or your
local distributor.
Empty the waste container and reset
usable volume of the waste container
as instructed by Chapter 5.3.1.
Turn off the power and contact Mindray
customer service department or your
local distributor.
Turn off the power and contact Mindray
customer service department or your
local distributor.
Enter the "Service → Status" screen
and record the "56V(V)" voltage;
Shut down the analyzer and contact
Mindray customer service department
or your local distributor.