Post-Cleaning Inspection
1. Before preparing for sterilization, all medical devices should be cleaned.
2. Generally, an un-magnified visual inspection using good light condition
is sufficient.
3. All parts should be checked for visible soil and/or corrosion. Particular
attention should be paid to:
• Mating surfaces
• Working channel entrances/exits
• Recessed features
Sterilizing the Ureteroscope
WARNING Do not sterilize the Semi-Rigid Ureteroscope using methods
Autoclave Steam Sterilization
Note Use Stryker Sterilization Tray (P/N 233-032-890) for Autoclave Steam
Sterilization.
1. Clean
a. Before sterilization, clean ureteroscope as indicated in the 'Cleaning'
section.
b. Ureteroscope must be dried prior to the next step.
2. Wrap
a. If ureteroscope is to be stored, double wrap in blue muslin paper or a
tyvek pouch.
b. If ureteroscope is sterilized in a sterilization tray, wrap the tray.
c. Please refer to "NOTE: Flash Sterilization" for unwrapped method.
3. Autoclave Parameter
a. The following autoclave parameters are the minimum time and
temperature parameters for sterilization:
Chamber
Configuration
Chamber
Temperature
Sterilization Time 4 minutes
Dry Time
Product
Configuration
other than those described in this manual. Infection or
device damage may result.
Flash
(Pre-Vac)
132°C
(270°F)
-
Unwrapped
Gravity
Pre-Vacuum
121°C
(250°F)
(270°F)
30 minutes
4 minutes
20 minutes
20 minutes
Double
Double
Wrapped
Wrapped
Gravity
132°C
132°C
(270°F)
15 minutes
60 minutes
Double
Wrapped
EN-12