Meter Setup Menu Features
The meter can also use manual salinity correction for dissolved oxygen readings
reported as mg/L . The manual salinity correction factor must be entered as ppt
(parts per thousand) .
Conductivity
Salinity
at 20 °C
Correction
(mS/cm)
Value (ppt)
5
3
6
4
7
4
8
5
9
6
10
6
11
7
12
8
13
8
14
9
15
10
16
10
17
11
18
12
19
13
This table was calculated from the International Oceanographic Tables, Vol . 1, National Institute of Oceanography
of Great Britain, Womley, Godaming, Surrey, England and Unesco, Paris 1971 .

Conductivity Setup Menu Features

Temperature Compensation and Reference Temperature
The Orion Star and Star Plus conductivity meters have the ability to use a
temperature compensation feature that calculates and displays the conductivity
measurements at a reference temperature of 5 °C, 10 °C, 15 °C, 20 °C or 25 °C
(Orion Star meters have a reference temperature of 15 °C, 20 °C or 25 °C only) .
The temperature compensation can be set as linear for most aqueous samples,
non-linear for ultra pure and low ionic strength samples or off for non-temperature
compensated conductivity measurements .
EN-66
Conductivity
at 20 °C
(mS/cm)
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
Thermo Scientific Orion Star
Salinity
Conductivity
Correction
at 20 °C
Value (ppt)
(mS/cm)
13
35
14
36
15
37
15
38
16
39
17
40
18
42
18
44
19
46
20
48
21
50
22
52
22
54
23
56
24
and Star Plus Series Meter User Guide
Salinity
Correction
Value (ppt)
25
25
26
27
28
29
30
32
33
35
37
38
40
42