Herunterladen Diese Seite drucken

Conductivity Temperature Display And Calibration - Thermo Scientific Orion Star series Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für Orion Star series:

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

Conductivity Technique
3 . Remove the conductivity probe from the sample, rinse it with distilled or
deionized water, blot it dry, place it in the next sample and repeat step 2 .
4 . Once all of the samples have been measured, rinse the conductivity probe with
distilled or deionized water and blot it dry . Consult the conductivity probe user
guide for proper storage techniques .

Conductivity Temperature Display and Calibration

Conductivity Temperature Display
Star Plus meters allow the temperature to be viewed on individual measurement
lines in addition to the temperature display on the top, left of the screen .
1 . In the measurement mode, press
arrow icon will point to the selected line .
2 . Press
/
be changed to display conductivity (µS/cm or mS/cm), total dissolved solids
(mg/L), salinity (ppt), resistivity (MΩ-cm), temperature (no icon) or a blank line .
Conductivity Temperature Calibration
The temperature calibration mode of the Star Plus meter allows the temperature on
each measurement line to be manually adjusted .
1 . In the measurement mode, press
and press
2 . Press
to begin the calibration .
3 . When the reading stabilizes, the arrow icon and the first digit will flash . Enter
the temperature by pressing
the next digit .
4 . Press
to save and end the calibration .
EN-42
to change the value on the selected line . The middle line can
/
until the temperature is shown for the selected line .
Thermo Scientific Orion Star
to choose the middle display line . The
to choose the middle measurement line
/
to adjust each digit and
and Star Plus Series Meter User Guide
to move to

Werbung