pH Troubleshooting
Problem:
The meter does not recognize the pH buffer value during calibration .
Solution:
Verify that the correct buffer set was selected in the setup menu . The
meter uses the raw mV reading of the electrode to recognize a buffer
during calibration . As the electrode ages or becomes dirty, its mV
readings will drift and you will need to manually enter the pH buffer value
when calibrating .

SE Troubleshooting

Problem:
It takes several minutes for the readings to stabilize during a calibration .
Solution:
The concentration range in the setup menu is set to low . Change the
concentration range to high .
The ISE resolution is set to 3 digits in the setup menu . Change the ISE
resolution to 2 digits for faster stabilization of the readings .
Problem:
When I use the automatic blank correction setting and calibrate an ISE,
the meter gives a slope that is too low or cannot be manually checked .
Solution:
Turn the automatic blank correction setting off in the setup menu .
Conductivity Troubleshooting
Problem:
The meter does not recognize the conductivity standard during
calibration .
Solution:
Verify that the default cell constant was entered in the setup menu . The
cell constant is usually printed on the conductivity probe cable . Verify
that the conductivity standard is one that is programmed into the meter .
Re-calibrate with a fresh standard .
Problem:
The temperature coefficient value does not appear in the setup menu .
Solution:
The current temperature compensation setting is nonlinear or off .
Change the temperature compensation to linear and the next screen will
be the temperature coefficient value entry screen .
Problem:
The measurement is out of range when it should be in range .
Solution:
Check that the conductivity probe is fully immersed in the solution .
Verify that the cell constant is correct for the conductivity probe that is
connected to the meter . Verify that the cell type selected in the setup
menu is set to Std .
Thermo Scientific Orion Star
and Star Plus Series Meter User Guide
Troubleshooting
EN-57