MP6025871-12C1 / A
Connecting the tester for testing or cleaning
Before proceeding, have available your pump user guide.
To connect the tester:
1
Hold the transmitter and the tester as shown. Line up the
flat side of the tester with the flat side of the transmitter.
2
Push the tester into the transmitter until the flexible side
arms of the tester click into the notches on both sides of the
transmitter.
3
Within five seconds, when properly connected, the green
light on the transmitter flashes for about 10 seconds.
4
To test the transmitter, check the sensor icon on the pump
to ensure that the transmitter is sending a signal (see your pump user guide).
5
To clean the transmitter, see Cleaning the transmitter, on page 9.
6
After testing or cleaning, disconnect the tester from the transmitter.
Disconnecting the tester
To disconnect the tester:
1
Hold the transmitter body as shown and pinch the side
arms of the tester.
2
With the tester arms pinched, gently pull the transmitter
away from the tester.
Note: To save transmitter battery life, do NOT leave the
tester connected after cleaning or testing.
Cleaning the transmitter
The transmitter is a single-patient use device and not intended for multi-patient use.
Caution:
Note: The tester is a required component for cleaning the transmitter. For details,
see Watertight Tester, on page 7.
Always clean the transmitter after each use.
To clean the transmitter, you will need the following materials: mild liquid soap, a soft-
bristled toddler toothbrush, a container, 70% isopropyl alcohol, and a few clean, dry
cloths.
Do not discard the transmitter in a medical waste container or
otherwise subject it to incineration. The transmitter contains a
battery that may explode upon incineration.
-9-