MP6025871-12C1 / A
Caution:
Only use the tester with the transmitter. Do not use any other test plug.
Inspecting the transmitter connector pins
This image is an example of how the connector pins should look.
Look inside the transmitter's connector opening to make sure that the connector pins are
not damaged or corroded. If the connector pins are damaged or corroded, the transmitter
cannot communicate with the charger or pump. Contact the 24 Hour HelpLine or your
local representative. It may be time to replace your transmitter.
Also look for moisture inside the connector opening. If you see any moisture, allow the
transmitter to dry for at least one hour. Moisture inside the connector opening could
cause the transmitter to not work properly, and could cause corrosion and damage over
time.
i
connector opening
connector pins
housing
-8-