Manual
Please note the data and test values specified in the leaflet
accompanying the calibration ampoules.
Break off the head of the ampoule. Take the calibration block in
your hand so that you can push in the calibration ampoule from the
bottom. Then place the calibration block on a level surface and
make sure that the calibration fluid runs into the calibration block.
Now insert the sensor head of the measuring instrument into the
calibration block carefully as far as it will go.
Wait two hours (acclimatisation time) before performing the
calibration according to the one-point calibration procedure
described under "CAL".
Remove the sensor head from the calibration block. Dispose of the
ampoule and clean the calibration block with distilled water.
Important: Only use the calibration ampoules once. The
temperature may not change during the acclimatisation time. Only
calibrate at room temperatures of 20 to 21 °C. Calibrations may
only be made with suitable reference values and by trained
personnel.