Calibration

Calibration of the climate sensors is only necessary in rare cases.
If high demands on the accuracy are made, we recommend an
annual onepoint calibration of the sensor.
You can perform one-point calibration yourself but we advise
against doing so because professional reference values are not
usually available.
Instead, make use of the possibility of certified calibration
according to DKD and ISO. Further information about this can be
obtained from your dealer.
One-point calibration (r.H.) with calibration device and
calibration ampoule:
Check the calibration device for foreign bodies or residue from
previous calibrations and clean the device if necessary.
Different calibration fluids are available for the moisture values 35
%, 50 % and 80 %. Only the 50 % calibration fluid should be used
for standard calibrations.