Herunterladen Diese Seite drucken

Ace instruments AFM-5 Bedienungsanleitung Seite 17

Atemalkoholtester
Vorschau ausblenden Andere Handbücher für AFM-5:

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
ACE INSTRUMENTS ADVISES
EVEN IF YOU GET A RESULT MEANING „SAFE RANGE", YOU
SHOULD NEVER DRIVE AFTER CONSUMING ALCOHOL OR
DO ANY OTHER ACTIVITIES THAT ARE SEVERELY IMPAIRED
BY ALCOHOL CONSUMPTION AND MIGHT ENDANGER YOUR
HEALTH AND SAFETY OR THE HEALTH AND SAFETY OF
OTHERS.
Calibration
To maintain a constant accuracy of the breathalyzer,
ACE Instruments recommends a calibration after 500 tests
or approx. every 12 months (depending which incident happens
first). For a calibration service you can send the device to
ACE Instruments (see www.alkomat.net). The calibration is not
mandatory, but the accuracy of the breathalyzer will decrease
steadily without it and additionally the warranty for the sensor
expires.
We recommend sending it as a parcel, so that in cases of
loss an inquiry or proof of delivery is possible.
Battery notices
The device associated with this operating manual contains
one (1) AAA Alkaline battery (1.5 V) in delivery condition. This
battery is not made for recharging. The use of a rechargeable
battery is not possible with this device and can damage it. If
the included battery is depleted, it must not be thrown into
the household waste. Used batteries may contain harmful
substances that can cause environmental or health damage.
Please return all depleted batteries to electrical appliance
stores, who collect them for free.
Battery exchange
To replace the battery, push the battery cover on the rear of
the unit up and replace the discharged battery with a new one.
Be aware of the polarity when inserting the battery (+/- pole).
Put the battery compartment cover on the rear panel and
press it gently on the device until it clicks. In the EU the return
of used batteries is mandatory by law. All depleted batteries
and also rechargeable batteries will be recycled. This way
valuable resources like iron, zinc and nickel can be retrieved
and are reusable then. Battery recycling is the easiest lesson
in environmental protection – so please support it!
23
The shown garbage bins mean that batteries or rechargeable
batteries must not be thrown into the household waste. The
signs below the garbage bins are acronyms for the chemical
substances inside the battery:
Technical data*
Sensor
Fuel Cell sensor
+/- 0.050‰ (per mill) with reference to
Accuracy
1.000‰ at 25°C (per mill = default setting)
OR +/- 0.005% BAC (Blood Alcohol Concen-
tration) with reference to 0.100% BAC at 25°C
OR +/- 0.025mg/l with reference to 0.500mg/l
at 25°C
Detection
0.000 up to 4.000‰ (per mill)
Range
In the App „Alcofind" you can change the
measurement unit into %BAC or mg/l.
Warm-up
Within 15 seconds at 0.5‰, it may vary
Time
depending on the value of per mill / BAC
which was measured
Connectivity
Bluetooth 4.0 BLE, Frequency band: 2402
up to 2480 MHz, Maximum transmit power:
-17.2 dBm
Dimension
68mm x 41.5mm x 16.9mm (L x W x H)
Weight
43g (battery included)
Power Supply
1 piece of 1.5V AAA size Alkaline battery
Working
5°C up to 40°C
Temperature
Storage
0°C up to 40°C
Temperature
Calibration
Every 12 months or after 500 tests.For a
calibration service you can send the device
to ACE Instruments (see www.alkomat.net).
* The technical data are subject to change without prior notice
for functional improvements.
Hg: Battery contains Mercury
Cd: Battery contains Cadmium
Pb: Battery contains Lead
24

Werbung