Standard delivery

Hi_Cal 150 mm
Hi_Cal 300 mm
Hi_Cal 450 mm
Including:
1 fixed probe, with saphire ball
1 master gauge
1 charger 9 V | 1100mA (mains AC voltage according to the country)
1 protection cover
1 calibration certificat
1 instruction manuel
Other accessories: see the site www.sylvac.ch

In the event of problems

Inactive corrections indicator :
Reactivate the corrections as follows
Important: This action must be done only with the base key (3mm ball)
Settings menu :
22
3 mm
1.
2.
Check the moving carriage alignment sensor :
Taking the reference
Lin correction, in µm/300mm
Test the UP/DOWN button; active 100%
Option to modify the motor release step to measure small
diameters with the Left/Right buttons.
Réf. N° : 830.0150
Réf N° : 830.0300
Réf N° : 830.0450
RS232 COR ON command
Retake the Ref
Switch the instrument on by pressing
ON + SETUP
Wait for the «init» display then browse
using the keys
initialisation.
Rest :
2.2 - 2.4 V
UP :
2.8 - 3.0 V
DOWN :
1.6 - 1.8
after instrument
V
4.0 - 4.1 V