Herunterladen Inhalt Inhalt Diese Seite drucken

Calibration And Adjustment - Thermo Scientific Finnpipette Bedienungsanleitung

Digital single channel multichannel
Vorschau ausblenden Andere Handbücher für Finnpipette:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 14
1. Depress the push button all the way to the second stop.
2. Dip the tip under the surface of the liquid in the
reservoir to a depth of about 1 cm, and slowly
release the push button.
This action will fill the tip. Withdraw the tip from
the liquid touching it against the edge of the
reservoir to remove excess liquid.
3. Deliver the preset volume by gently depressing the
push button to the first stop.
Hold the push button at the first stop. Some liquid
will remain in the tip and this should not be
included in the delivery.
4. The remaining liquid should either be discarded with the tip or pipetted back into the container.
REPETITIVE TECHNIQUE
The repetitive technique offers a rapid and simple procedure for repeated delivery of the same
volume. Fill a clean reagent reservoir with the liquid to be dispensed.
1. Depress the push button all the way to the second stop.
2. Dip the tip under the surface of the liquid in the
reservoir to a depth of about 1 cm, and slowly
release the push button. This action will fill the tip.
Withdraw the tipfrom the liquid touching against
the edge of the reservoir to remove excess liquid.
3. Deliver the preset volume by gently depressing the
push button to the first stop.
Hold the push button at the first stop. Some liquid
will remain in the tip and this should not be included
in the delivery.
4. Continue pipetting by repeating steps 2 and 3.
PIPETTING OF HETEROGENEOUS SAMPLES
(deproteinization in blood glucose determination, for example)
Use steps 1 and 2 of the forward technique to fill the tip with blood.
Wipe the tip carefully with a dry clean tissue.
1. Immerse the tip into the reagent and depress the
push button to the first stop, making sure the tip is
well below the surface.
2. Release the push button slowly to the ready
position. This will fill the tip. Keep the tip in the
solution.
3. Depress the push button to the first stop and
release slowly. Keep repeating this procedure until
the interior wall of the tip is clear.
4. Finally, depress the push button all the way to the second stop to completely empty the tip.

CALIBRATION AND ADJUSTMENT

All Finnpipettes are factory calibrated and adjusted to give the volumes as specified with
distilled or deionized water using the forward pipetting technique. It should be noted that the
use of other pipetting techniques may affect the calibration results. The pipettes are constructed
to permit re-adjustment for other pipetting techniques or liquids of different temperature and
viscosity.
DEVICE REQUIREMENTS AND TEST CONDITIONS
An analytical balance must be used. The scale graduation value of the balance should be chosen
according to the selected test volume of the pipette:
Volume range
under 10 µl
10-100 µl
above 100 µl
readable graduation
0.00 1 mg
0.01 mg
0.1 mg
A
1 2 3 4 5
B
C
A
1 2 3 4
. . .
B
C
A
1 2 3 4
. . .
B
C
7

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis