STERILIZATION
The entire pipette can be sterilized by autoclaving it at 121°C (252°F) (minimum 20 minutes). No
special preparations are needed for autoclaving. you can use steam sterilization bags if needed.
After autoclaving the pipette must be cooled to room temperature for at least two hours. Before
pipetting, make sure that the pipette is dry. We recommend that you check the calibration of
0.5-1000 µl pipettes after every 25th sterilization cycle and of 1-10 ml pipettes after every 10th
sterilization cycle.
The table below lists possible problems and their solutions.
Defect
Possible reason
Leakage
Tip incorrectly attached
Foreign particles between tip and
tip cone
Foreign particles between the
piston, the O-ring and the cylinder
Insufficient amount of grease on
cylinder and O-ring
O-ring damaged
Incorrect operation
Inaccurate
dispensing
Tip incorrectly attached
Calibration altered: caused by
misuse, for example
Piston jammed
Pipette has been unused for a
long period
Highly volatile solvent is pipetted
repeatedly cleaning the grease
Inaccurate
Unsuitable calibration
dispensing with
High viscosity liquids may require
certain liquids
recalibration
CAUTION!
The Finnpipette is designed to allow easy in-lab service. If you would prefer to
have us or your local representative service your pipette, please make sure that
the pipette has been decontaminated before you send it to us.
Please note that the postal authorities in your country may prohibit or restrict the
shipment of contaminated material by mail.

TROUBLE SHOOTING

Solution
Attach firmly
Clean tip cones attach new tips
Clean and grease O-ring and
cylinder.
Grease accordingly
Change the O-ring
Follow instructions carefully
Attach firmly
Recalibrate according to
instructions
Remove tip if attached. Press
the plunger to the second stop
and release several times to re-
spread the grease
Recalibrate with the liquids
in question
13