Herunterladen Inhalt Inhalt Diese Seite drucken

Classification Output 5 To 40 Classes Within Tolerances (Tt60) - TESA TT20 Gebrauchsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Command R-M «Sink»
+5V
6
14
R-M
Retour
7
GND/R-M
GND
12
5 ÷ 48V
14
R-M
Retour
7
GND/R-M

8.4 Classification output 5 to 40 classes within tolerances (TT60)

Additional to the optics classification system «Within tolerance» (green), «Rework» (yel-
low), «Reject» (red) with the signals to the internal relays, the TT60 proposes an additio-
nal choice of 5, 10, 20 or 40 classes plus «Rework» and «Reject». The selection is done by
means of the configurations switches 6 and 7 on the back side of the instrument.
These X classes represent always the tolerance range defined by the lower limit and the
upper limit divided in X classes having same band width.
The functioning of the X classes output depends the tolerance limits setting.
The X classes output is given by a synchronous output on the 15-pin female connector.
This synchronous output is parallel active to the classification by LEDs and by internal
relays.
8.4.1 Form of output signals
Sub-D 15 pf
Clock - 3
Data - 2
Strobe - 1
PC354
PC354
Command R-M «Source»
+5V
6
R-M
14
Retour
7
GND/R-M
GND
12
5 ÷ 48V
R-M
14
GND
Retour
7
GND/R-M
0.1 ms
PC354
PC354
4.8 KHz
0.1 ms
75

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Tt60

Inhaltsverzeichnis