Problem
Pulse oximeter
shows inter-
rupted meas-
urements or
severe jumps in
the measure-
ment.
Unable to
transfer the
measurements.
174 GB/MT
Possible cause
Circulation problems in the
finger being measured.
The finger being measured is
too large or too small.
The finger, hand or body are
moving.
Arrhythmia.
The "HealthForYou" app is not
activated or Bluetooth
in the smartphone settings.
The batteries in the pulse
oximeter are low or drained.
The device is still on.
solution
Refer to warnings and safety notes.
The fingertip must be the following
size: between 10 and 20 mm wide.
Between 5 and 15 mm thick.
Keep the finger, hand and body still
during measurement.
Seek medical attention.
Activate Bluetooth
is off
smartphone and start the app.
®
Replace the batteries.
Wait for the device to switch off.