Viatom Checkme Lite Benutzerhandbuch Seite 17

Health monitor
Vorschau ausblenden Andere Handbücher für Checkme Lite:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

<5 % UT
(>95 % dip in UT)
for 5 s
Power
3 A/m
frequency
(50/60
HZ)
magnetic field
IEC 61000-4-8
Note: U
is the AC mains voltage prior to application of the test level.
T
Guidance and Declaration - Electromagnetic Immunity
The Health Monitor is intended for use in the specified electromagnetic environment. The
customer or the user of the Health Monitor should assure that it is used in such an environment
as described below.
Immu
IEC60601
nity
test level
test
Condu
3 Vrms 150
ced RF
kHz to
IEC61
80 MHz
000-4-
outside ISM
6
bands
Radiat
3 V/m 80
ed RF
MHz to
IEC61
2.5 GHz
000-4-
3
Note 1: At 80 MHz to 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
a
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
<5 % UT
(>95 % dip in UT)
for 5 s
3 A/m
Compliance
Electromagnetic environment - guidance
level
3 Vrms 150
Portable and mobile RF communications equipment
kHz to
should be used no closer to any part of the system,
80 MHz
including cables, than the recommended separation
outside ISM
distance calculated from the equation appropriate for
bands
the frequency of the transmitter. Recommended
separation distances:
3 V/m 80
Recommended separation distances:
MHz to
80 MHz~800 MHz:
2.5 GHz
800MHz-2.5GHz:
Where, P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
should be less than the compliance level in each
frequency range
Interference may occur in the vicinity of equipment
marked with the following symbol:
a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
d
d
=
d
.
b
15
=
1
2 .
P
=
1
2 .
P
2
3 .
P
,
a