RF emissions CISPR 11
RF emissions CISPR 11
Harmonic
emissions
IEC61000-3-2
Voltage Fluctuations /
Flicker Emissions IEC
61000-3-3
Guidance and Declaration - Electromagnetic Immunity
The Health Monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the Health Monitor should assure that it is used in such an environment.
Immunity test
IEC60601 test level
Electrostatic
± 6 kV contact
discharge
± 8 kV air
(ESD)
IEC
61000-4-2
Electrical fast
± 2 kV for power
transient/burs
supply lines
t
±
IEC 61000-4-4
input/output lines
Surge
± 1 kV line(s) to
IEC 61000-4-5
line(s)
± 2 kV line(s) to
earth
Voltage dips,
<5 % UT
short
(>95 % dip in UT)
Interruptions
for 0.5 cycle
and
Voltage
40 % UT
variations on
(60 % dip in UT)
power supply
for 5 cycles
input lines
70 % UT
IEC 61000-4-
(30 % dip in UT)
11
for 25 cycles
Group 1
Class B
Class A
Complies
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
1
kV
for
±
input/output
lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
The device uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
The
device
is
establishments,
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Electromagnetic environment -
guidance
Floors should be wood,
concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at
least 30 %.
Mains power quality should be
that of a typical commercial or
1
kV
for
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of our product requires
continued operation during
power mains interruptions, it is
recommended that our product
be powered from an
uninterruptible power supply or
14