Herunterladen Inhalt Inhalt Diese Seite drucken

Checking Functions - Mettler Toledo InPro 6800 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

10
2)
Polarization time:
Time span during which the sensor is under a polarization volt-
age.
Please note that a sensor which is not properly polarized will give
results which will not be accurate. A sensor which has been polar-
ized for a too short period of time may give an accuracy of ±30%.
2.2.3

Checking functions

Battery check
InPro 6800/6900/6950
52 200 952
By pushing the switch to the left to the posi-
tion "1 – Battery check", the O
ter will control the polarization voltage. If the
green LED lights up than the battery is still ok
and delivers the correct polarisation voltage. If
not, the battery must be changed (see section
2.3). If after a battery change, the LED does
not light up you will have to return it to your
METTLER TOLEDO dealer.
© 03/14 Mettler-Toledo AG, Greifensee
Accessories O
2
Sensor-Mas-
2
Printed in Switzerland
sensors

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Inpro 6900Inpro 6950

Inhaltsverzeichnis