Manual
Important: calibrations should be
carried out only by skilled people using
an appropriate calibration equipment.
CAL2: CAL2 (one point calibration) gives the user the possibility to
set an offset for sensor 2 (rel. humidity). The offset will base on the
lower adjustment point (11% r.h.). The adjustment point should lay
between 30% ... 95% r.h. (see figure).
The offset value appears on the top of the display. Max. +/- 10%rH
can be set. CAL2 is to be selected in connection with the unity for
rel. humidity (%) only.
11%
Factory calibration can be obtained by entering offset 0.0
HUMIPORT 10/20 english
Factory calibration
Offset
New calibration
line
Measured value
page 28