For the purposes of EN 60601 -1-2 the Smoke C heck monitor has an essential
pe rformance (the Smoke Check should continue to indicate a le vel of 11-20
p.p.m. during a te st, whilst measuring gas from a 15 p.p.m. CO gas bottle)
Guidance and Manufacturer's Declaration – Electromagnetic Emission
The Sm oke C heck is intended for use in the electromagnetic e nvironment specified
be low. It is the responsibility of the customer or the user of the Smoke C heck should
e nsure that it is used in such an environment
Emission Test
R F e m issions
C ISPR 11
R F e m issions
C ISPR 11
Harm onic emissions
IEC 61000-3-2
Voltage fluctuations /
flick e r e missions
IEC 61000-3-3
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Sm oke C heck is intended for use in the electromagnetic e nvironment specified
be low. It is the responsibility of the customer or the user of the Smoke C heck should
e nsure that it is used in such an environment
Immunity Test
Ele ctrostatic
discharge (ESD)
IEC 61000-4-2
Ele ctrical fast
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Compliance
Group 1
Group B
Not Applicable
(batte ry powe red)
Not Applicable
(batte ry powe red)
IEC 60601
test level
± 6 k V contact
± 8 k V air
± 2 k V for powe r
supply lines
± 1 k V for input /
output lines
± 1 k V line (s)
to line (s)
± 2 k V line (s)
to e arth
Electromagnetic Environment -
Guidance
The Sm oke C heck uses RF energy
only for its internal function. Therefore,
its R F e mission is very low and is not
lik e ly to cause any interference in
ne arby electronic e quipment.
The Sm oke C heck is suitable for use in
all e stablishments, including domestic
and those directly connected to the
public low-voltage powe r supply
ne twork that supplies buildings used
for domestic purposes
Compliance
Electromagnetic
Level
Environment - Guidance
± 6 k V contact
Floors should be wood,
± 8 k V air
concre te or ceramic tile.
If floors are covered with
synthe tic m aterial, the
re lative humidity s hould
be at le ast 30%.
Not Applicable
Mains powe r quality
(Batte ry
should be that of a typical
powe re d)
com m ercial or hospital
and no cables
e nvironment.
Not Applicable
Mains powe r quality should
(Batte ry
be that of a typical
Powe re d) and
com m ercial or hospital
no cables.
e nvironment.
59