For the purposes of EN 60601 -1-2 the Smoke C heck monitor has an essential
pe rformance (the Smoke Check should continue to indicate a le vel of 11-20
p.p.m. during a te st, whilst measuring gas from a 15p.p.m. C O gas bottle)
Guidance and Manufacturer's Declaration – Electromagnetic Emission
The Sm oke C heck is intended for use in the electromagnetic e nvironment specified
be low. It is the responsibility of the customer or the user of the Smoke C heck should
e nsure that it is used in such an environment
Emission Test
R F e m issions
C ISPR 11
R F e m issions
C ISPR 11
Harm onic emissions
IEC 61000-3-2
Voltage fluctuatuins /
flick e r e missions
IEC 61000-3-3
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Sm oke C heck is intended for use in the electromagnetic e nvironment specified
be low. It is the responsibility of the customer or the user of the Smoke C heck should
e nsure that it is used in such an environment
Immunity Test
Ele ctrostatic
discharge (ESD)
IEC 61000-4-2
Ele ctrical fast
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Compliance
Group 1
Group B
Not Applicable
(batte ry powe red)
Not Applicable
(batte ry powe red)
IEC 60601 test
level
± 6 k V contact
± 8 k V air
± 2 k V for powe r
supply lines
± 1 k V for input /
output lines
± 1 k V line (s) to
line (s)
± 2 k V line (s) to
e arth
Electromagnetic Enviroment -
Guidance
The Sm oke C heck uses RF energy only
for its inte rnal function. Therefore, its
R F e m ission is very low and is not likely
to cause any interference in nearby
e le ctronic e quipment.
The Sm oke C heck is suitable for use in
all e stablishments, including domestic
and those directly connected to the
public low-voltage powe r supply network
that supplies buildings used for domestic
purposes
Compliance
Electromagnetic
Level
Enviroment -Guidance
± 6 k V contact
Floors should be wood,
concre te or ceramic tile.
± 8 k V air
If floors are covered with
synthe tic m aterial, the
re lative humidity should
be at le ast 30%.
Not Applicable
Mains powe r quality
should be that of a typical
(Batte ry
com m ercial or hospital
powe re d)and
e nvironment.
no cables
Not Applicable
Mains powe r quality should
be that of a typical
(Batte ry
com m ercial or hospital
Powe re d) and
e nvironment.
no cables.