Recommended separation distances between
portable and mobile RF communications equipment and the Smoke Check
The Sm oke C heck is intended for use in an electromagnetic e nvironment in which
radiated R F disturbance s are controlled. The customer or the user of the Smoke Check
can he lp prevent e lectromagnetic interfere nce by maintai ning a minimum distance
be twe e n portable and mobile RF communications equipment (transmitters) and the
Sm ok e Check as re commended below, according to the maximum output powe r of the
com m unications equipment
R ate d Maximum
O utput Powe r of
Transmitter in
W atts (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output powe r not listed above, the
re com mended separation distance d in m etres (m) can be e stimated using the
e quation applicable to the fre quency of the transmitter, whe re P is the maximum
output powe r rating of the transmitter in watts (W) according to the transmitter
m anufacture r.
NO TE 1
At 80 MHz and 800 MHz, the separation distance for the higher fre quency
range applies.
NO TE 2
The se guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and re flection from structures obje cts and
pe ople
.
Se paration Distance in Meters (m) according to Frequency of
Transmitter
150 KHz to 80 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
80 MHz to 800 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
30
800 MHz to 2.5 GHz
d = 2.3√P
0.23
0.73
2.3
7.3
23.3