The C O reading will rise to a plateau over the course of several
seconds. The final value will be held until the unit is turned off and
will be displayed in one of the four ranges:
0–6 ppm with green indicator
7–10 ppm with amber indicator
11–20 ppm with red indicator
20+ ppm with flashing red indicator and alarm
Important note:
Before repeating a measurement the unit must be turned off,
and the mouthpiece and adapter removed for at least 1 minute.
This is to allow re-equilibration with ambient air and to dry the
surface of the sensor. Visually inspect that a ll moisture has
evaporated from the surface of the sensor before reuse.
If the unit is switched on again too quickly after use there may
be a response to residual expired carbon monoxide from the
previous test.
In this case the display will show:
Blow
0-6 ppm
Gas
17