Herunterladen Inhalt Inhalt Diese Seite drucken

General Information - Hach Polymetron 9500 Benutzerhandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 26
Expanded manual version
For additional information, refer to the expanded version of this manual, which is available on the
manufacturer's website.
Specifications
Specifications are subject to change without notice.
Specification
Component description
Operating temperature
Storage temperature
Enclosure
European standards
ETL approved
Power requirements
Altitude requirements
Pollution degree /
Installation category
Outputs
Relays
Dimensions
Weight
Digital communication
Data logging
Warranty

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer's website.
Details
Microprocessor-controlled and menu-driven controller that operates the sensor and
displays measured values.
-20 to 60 ºC (-4 to 140 ºF); 95% relative humidity, non-condensing with sensor load
<7 W; -20 to 50 ºC (-4 to 104 ºF) with sensor load <28 W
-20 to 70 ºC (-4 to 158 ºF); 95% relative humidity, non-condensing
NEMA 4X/IP66 metal enclosure with a corrosion-resistant finish
EN 61326-1: 2006; EN 61010-1:2010
File 65454
AC powered controller: 100-240 VAC ±10%, 50/60 Hz; Power 50 VA with 7 W
sensor/network module load, 100 VA with 28 W sensor/network module load
(optional Modbus RS232/RS485 or Profibus DPV1 network connection).
24 VDC powered controller: 24 VDC—15%, + 20%; Power 15 W with 7 W
sensor/network module load, 40 W with 28 W sensor/network module load (optional
Modbus RS232/RS485 or Profibus DPV1 network connection).
Standard 2000 m (6562 ft) ASL (Above Sea Level)
Pollution Degree 2; Installation Category II
Two analog (0-20 mA or 4-20 mA) outputs. Each analog output can be set to
0-20 mA or 4-20 mA, and can be assigned to represent a measured parameter such
as pH, temperature, flow or calculated values. Optional 3 additional analog outputs.
Secure Digital Memory card for use in data logging and software updates.
Four SPDT, user-configured contacts, rated 5A 250 VAC (resistive). Contacts are
rated 250 VAC, 5 Amp resistive maximum for the AC powered controller and 24 VDC,
5A resistive maximum for the DC powered controller. Relays are designed for
connection to AC Mains circuits (i.e., whenever the controller is operated with 115 -
240 VAC power) or DC circuits (i.e., whenever the controller is operated with 24 VDC
power).
½ DIN—144 x 144 x 180.9 mm (5.7 x 5.7 x 7.12 in.)
1.7 kg (3.75 lb)
Optional Modbus RS485/RS232 or Profibus DPV1 network connection for data
transmission
Secure Digital Card or special RS232 cable connector for data logging and
performing software updates
2 years
English 3

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis