• Do not attempt any maintenance other than the possible solutions listed
below.
Problem
LifeStyle does not
operate when the flow
selector is NOT in the
OFF position.
A continuous alarm
sounds and the (red)
ALARM light illuminates
continuously.
Intermittent alarm
condition (1-second
warning sounds every 5
seconds), and the
(yellow) BATTERY light
illuminates
intermittently.
Rapid alarm condition
(½-second alarm
sounds every ½
second), and the
(yellow) BATTERY light
illuminates
intermittently.
14-GB
Probable Cause
Battery is discharged.
Battery is not installed
correctly.
Malfunction.
LifeStyle has not
detected a breath for a
predetermined period
of time.
Battery requires
charging.
Battery voltage is too
low to operate the
LifeStyle unit.
LifeStyle
Solution
Power the unit through the
DC power adapter/socket,
or an AC adapter/socket, or
replace with a fully charged
battery.
Remove the battery and
reinsert it, ensuring that it is
in the correct position.
Contact your Equipment
Provider, and change to a
reserve oxygen supply.
Check the cannula
connection.
Ensure that cannula tubing
is not kinked.
Make sure that the cannula
is positioned properly and
that you are breathing
through your nose.
Replace battery or connect
to a DC or an AC socket
within 5 minutes.
Replace battery or connect
to a DC or an AC socket
immediately.