Status
Audible Alarm Light
Intermittent:
Warning
Beep, beep,
beep, pause,
beep, beep,
beep...
Continuous:
Alarm
Beep
Intermittent:
Alarm
Beep, beep,
beep...
Rapid
Alarm
intermittent:
Beep, beep,
beep...
Rapid
Alarm
intermittent:
Beep, beep,
beep...
AirSep
Corporation
®
Indicates
(Red) ALARM;
Breathing rate
intermittent
i s
light
approaching
the threshold
of LifeStyle's
capacity; or
the LifeStyle
unit is
approaching
general
malfunction.
(Red) ALARM;
No breath
continuous
detected by
light
the unit for a
predeter-
mined time
period.
(Yellow)
Battery
BATTERY;
voltage is too
intermittent
low to operate
light
LifeStyle.
(Red) ALARM;
Breathing rate
intermittent
is exceeding
light
the capacity of
the LifeStyle
unit.
(Red) ALARM;
General
continuous
malfunction of
light
the LifeStyle
unit has
occurred.
GB
Action
Reduce activity, and/
or locate another
source of oxygen.
LifeStyle can still
function during this
condition.
Check the cannula
connection. Ensure
that you are
breathing through
your nose. If the
alarm persists,
contact your
Equipment Provider.
Replace the battery,
or connect the
LifeStyle unit into a
DC vehicle cigarette
lighter socket or an
AC socket
immediately.
Reduce activity, and/
or locate and use
another source of
oxygen. Contact your
Equipment Provider.
Turn off the unit.
Change to another
source of oxygen,
and contact your
Equipment Provider.