HEIDENHAIN EIB 741 Installation/Inbetriebnahme Seite 30

12/2010
Vorschau ausblenden Andere Handbücher für EIB 741:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Electrical Connection
Note:
GND pin from X31 or X32 is connected to GND in the unit, i.e. with the encoder connections as well.
A trigger is triggered at each edge from LOW to HIGH. In reference to the differential signals this means:
U
: Low to High
Trigger,In+
U
: High to Low
Trigger,In–
Recommendation: The external trigger source should address the EIB 741 via differential signals.
If the application requires the use of single-ended signals, the following applies:
The sensitivity to interference increases, since the EMC guidelines under which the EIB 741 was tested are not maintained. The
use of 5 V TTL signals is recommended.
The differential signals should then be wired as follows:
The vacant input should be kept open and the terminating resistor must be switched off.
Use of U
: Trigger upon positive edges
Trigger,In+
Use of U
: Trigger upon negative edges
Trigger,In–
Electrical properties:
Parameters
Input voltage range
Input resistor
30
U
To GND
Trigger,In
U
-
Terminal resistor out
Trigger,In+
U
Trigger,In–
U
-
Terminal resistor in
Trigger,In+
U
Trigger,In–
In+ to In–
Terminal resistor out
In+ to In–
Terminal resistor in
Min.
Max.
–2 V
+10.0 V
–10.0 V
+10.0 V
–5.5 V
+5.5 V
12.0 k−
105 −
160 −