Test
(Switch Position)
K
AC Volts
Min/Max (337)
L
DC Volts
R
Continuity
e
Ohms
?
Amps AC
*INRUSH
A
DC Amps
Frequency (337)
*The "INRUSH" button must be pressed after current is applied to the 50-turn coil.
**Display should bar (- - - -), then show reading with inrush light on (334, 335)
Table 2. Performance Tests 336, 337 (cont.)
5520A Output
20 V 50 Hz
600 V 50 Hz
600 V 400 Hz
600 V 400 Hz
20 V 400 Hz
20 V
-20 V
600 V
-600 V
25 Ω
600 Ω
6 KΩ
20 A 50 Hz
200 A 50 Hz
200 A 400 Hz
600 A 50 Hz
900 A 50 Hz (337 only)
**5 A 60 Hz
0.0 A after button push
20.0 A
-20.0 A
600.0 A
-600.0 A
900.0 A (337)
-900.0 A (337)
5 A 10 Hz
10 A 300 Hz
Meter Response
Lower Limit
19.3 V AC
593.5 V AC
563.5 V AC
563.5 V AC
Reading does not change
19.3 V DC
-20.7 V DC
593.5 V DC
-606.5 V DC
Beeper On
590.5 Ω
5905 Ω
19.1 A AC
195.5 A AC
187.5 A AC
587.5 A AC
881.5 A AC
4.4 A AC
-0.5 A DC
19.1 A DC
-20.9 A DC
587.5 A DC
-612.5 A DC
881.5 A DC
-918.5 A DC
9.4 Hz
298.0 Hz
ClampMeters
Performance Tests
Upper Limit
20.7 V AC
606.5 V AC
636.5 V AC
636.5 V AC
20.7 V DC
-19.3 V DC
606.5 V DC
-593.5 V DC
609.5 Ω
6095 Ω
20.9 A AC
204.5 A AC
212.5 A AC
612.5 A AC
918.5 A AC
5.6 A AC
0.5 A DC
20.9 A DC
-19.1 A DC
612.5 A DC
-587.5 A DC
918.5 A DC
-881.5 A DC
10.6 Hz
302.0 Hz