5.2 Error messages

During the program course an acoustic warning signal can be heard and the displayed
vacuum value is marked by an asterisk.
The set value is not achieved. Check whether leakage of the cover exists and for this purpose -
press the mixing bowl against the coupling once more. If the same error occurs during the use of a
mixing bowl / cover, the mixing bowl / cover must be exchanged.
Check if the seal in the coupling is clean and undamaged.
After the end of the program the error message "Error vacuum" is displayed.
During the stirring process the set vacuum has not been achieved. The stirring result may be
unsatisfactory. Confirm the error message by pressing the <Start/Stop> key.
After the end of the program the error message "Filter!" is displayed.
The filter element is clogged. Change the filter element as described in section 4.2.
During stirring operation the error message "Error overload" is displayed and the stirring
mechanism stops.
The motor of the stirring mechanism is overloaded. Check smooth running of the stirring
mechanism and the drive system.