Use according to specification

The machine presented here is a medical product (laboratory centrifuge) according to the IVD guideline 98/79/EG.
The centrifuge is used to separate substances or substance mixtures with a density of max. 1.2 kg/dm³. This also
includes substances and substance mixtures of human origin. The centrifuge is only intended to be used for this
purpose. A different use or application over and above this is deemed not in accordance with the specifications. The
company Andreas Hettich GmbH & Co. KG undertakes no liability for damages resulting therefrom.
Belonging to the application according to specification is also the observance of all references contained in the
Instruction Manual and compliance with the inspection and maintenance works.
2

Residual risks

The machine is constructed according to the state of the art and the recognized technical safety regulations.
Improper use and handling can result in dangers to life and limb of the user or third parties and impairments to the
machine or to other material assets. The machine is only to be used for the specified applications and only in an
impeccable technical safety condition.
Disturbances that can interfere with the safety are to be immediately rectified.
3

Technical specifications

Manufacturer
Model
Type
Mains voltage (± 10%)
Mains frequency
Connected load
Current consumption
Max. capacity
Allowed density
Speed (RPM)
Force (RCF)
Kinetic energy
Obligatory inspection (BGR 261)
Ambient conditions (EN 61010-1)
− Set-up site
− Altitude
− Ambient temperature
− Humidity
− Excess-voltage category
(IEC 60364-4-443)
− Pollution degree
Device protection class
EMC
− Emitted interference (suppression
of radio interference)
− Interference immunity
Noise level (dependent on rotor)
Dimensions
− Width
− Depth
− Height
Weight
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
ROTOFIX 32 A
1206
208 – 240 V 1∼
50 – 60 Hz
300 VA
1.4 A
4 x 100 ml / 32 x 15 ml
Up to 2000 m above sea level
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
Not suitable for use in explosion-endangered areas.
EN 55011,
Group 1, Class B
EN 61000-3-2
EN 61000-3-3
EN 61000-6-2
1206-01
100 – 127 V 1∼
50 – 60 Hz
300 VA
3.0 A
3
1.2 kg/dm
6000
4186
3160 Nm
no
Indoors only
2°C to 40°C
ΙΙ
2
Ι
FCC Class B
----
≤ 57 dB(A)
366 mm
430 mm
257 mm
23 kg
EN
19/72