Herunterladen Inhalt Inhalt Diese Seite drucken

Rotors And Attachments; Trunnions; Rotors And Accessories With Limited Term Of Use; Autoclaving - Hettich ROTOFIX 32 A Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für ROTOFIX 32 A:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

20.2

Rotors and Attachments

In order to prevent corrosion and material changes, rotors and accessories must be cleaned regularly with soap
or a mild detergent and a damp cloth. Cleaning is recommended at least once a week, even better after every
usage.
If the rotor or accessory parts are contaminated by pathogenic or radioactive material, a suitable cleaning has to
be executed. For disinfection we recommend Helipur
removal of radioactive material we recommend decon neutracon
Limited.
The rotors and accessory parts must be dried immediately after cleaning.
Angle rotors, container and hanger made of aluminium are to be lightly greased after drying using acid-free
grease, e.g. vaseline.
With aerosol-sealed rotors and bio safety systems (see Chapter "Anhang/Appendix, Rotoren und
Zubehör/Rotors and accessories") the sealing rings are to be checked and cleaned regularly (weekly). The
sealing ring is to be replaced immediately upon indication of crack formation, embrittlement or abrasive wear. To
prevent the sealing ring from twisting when the lid is opened and closed, the sealing ring must be lightly greased
with acid-free grease, such as Vaseline.
In order to prevent corrosion as a result of moisture between the rotor and the motor shaft, the rotor should be
disassembled and cleaned at least once a month, and the motor shaft should be lightly greased.
The rotors and accessory parts are to be checked on a monthly basis for corrosion damage.
Rotors and attachments may no longer be utilised upon indication of wear and tear or corrosion.
Check the firm seating of the rotor on a weekly basis.
20.2.1

Trunnions

With swing-out rotors the trunnions must be regularly lubricated (Hettich Lubricating Grease No. 4051) in order to
ensure consistent swinging out of the hangers.
20.2.2

Rotors and accessories with limited term of use

The use of specific rotors, suspensions and accessories is time limited.
They are marked with an expiry date, e.g. "einsetzbar bis Ende: / usable until end of: Ι V. Quartal 2011" (applicable
until the end of: Ι Vth quarter 2011).
The rotors, suspensions and accessories may not be used for longer periods for safety reasons once the
marked expiry date has been reached.
20.3

Autoclaving

Swing-out rotors, angle rotors made of aluminium, suspension made of metal, lids with biodegradable seals as well
as stands and reductions can be autoclaved at 121° C / 250°F (20 mins).
Otherwise you must ask the manufacturer.
The lids of the rotors and containers must be removed prior to autoclaving.
Autoclaving accelerates the ageing process of plastics. In addition, autoclaving may discolour plastics.
After autoclaving, we recommend that the sealing rings of the aerosol-tight and bio-safety systems be
exchanged.
20.4

Centrifuge containers

With leakiness or after the breakage of centrifuging containers broken container parts and leaked centrifugation
material are to be completely removed.
The rubber inserts as well as the plastic sleeves of the rotors are to be replaced after a glass breakage.
Remaining glass splitters cause further glass breakage!
If this concerns infectious material, a disinfection process is to be executed immediately.
®
H plus N from the company B. Braun Melsungen. For the
®
from the company Decon Laboratories
EN
27/72

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis