Dismounting
Fault clearance
The first measures to be taken are to check the
output signals as well as to evaluate the error
messages via the indicating and adjustment
module. The procedure is described below. Further
comprehensive diagnostics can be carried out on a
PC with the software PACTware™ and the suitable
DTM. In many cases, the causes can be deter-
mined in this way and faults can be rectified.
24 hour service hotline
However, should this measures not be successful,
call the VEGA service hotline in urgent cases under
the phone no. +49 1805 858550.
The hotline is available to you 7 days a week
round-the-clock. Since we offer this service world-
wide, the support is only available in the English
language. The service is free of charge, only the
standard telephone costs will be charged.
Fault messages via the indicating/adjustment
module
?
E013
l
no measured value available
à
sensor in boot phase
à
sensor does not find an echo, e.g. because
of faulty installation or incorrect parameter
adjustment
?
E017
l
Adjustment span too small
à
Carry out a fresh adjustment and increase
the distance between min. and max.
adjustment
?
E036
l
no operable sensor software
à
Carry out a software update or send the
instrument for repair
?
E041
l
Hardware error, electronics defective
à
Exchange instrument or return instrument
for repair
36
8 Dismounting

8.1 Dismounting steps

Warning:
Before dismounting, be aware of dangerous
process conditions such as e.g. pressure in
the vessel, high temperatures, corrosive or
toxic products etc.
Take note of chapters "Mounting" and "Connecting
to power supply" and carry out the listed steps in
reverse order.
8.2 Disposal
The instrument consists of materials which can be
recycled by specialised recycling companies. We
use recyclable materials and have designed the
electronic modules to be easily separable.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE
directive 2002/96/EG and the respective national
laws (in Germany, e.g. ElektroG). Pass the instru-
ment directly on to a specialised recycling com-
pany and do not use the municipal collecting
points. These may be used only for privately used
products according to the WEEE directive.
Correct disposal avoids negative effects to persons
and environment and ensures recycling of useful
raw materials.
Materials: see chapter "Technical data"
If you cannot dispose of the instrument properly,
please contact us about disposal methods or
return.