A special adapter cable is required for connection (order no.
087H0121). The coloured wires are to be connected as follows:
6.3 Function module pulse input
Module for two additional counters.
The pulse input 1 is marked as "I1 - | _ ", the input 2 as "I2 - | _ "
· Pulse inputs are programmable (IZAR@SET) for a value of:
1, 2.5, 10, 25, 100, 250, 1000 or 2500 litres per pulse.
· Possible units are all the energy units available in the meter, the
volume unit m³ the range < _ 8Hz; minimum pulse duration 10ms
· Input resistance 2.2MΩ; terminal voltage 3V DC
· Data is accumulated separately in registers
· Indicated in display as IN1 and IN2 and also to be transferred
over the communication modules
· Cable length up to 10 m
English
62 = brown;
63 = white;
64 = green