LEDs on Controller 507
Red LED „Too high" is lit = The safety sensor (3) is
active or has had contact with water. The metering
pump is switched off immediately, and a warning
signal is sounded.
Red LED „Too high" and red LED „Too low" are
lit and are flashing simultaneously = Water has
penetrated into the controller and has switched off
all functions. Open the device (see overleaf); dry the
PCB and clean, if and when necessary. If both LEDs
light up again after some hours, residual salts and
high humidity may be the cause. The device has to
be examined in a specialised workshop.
Controller 507 et diodes LED
LED rouge «Too high» est allumée = le capteur de
sécurité (3) est actif et en contact avec l'eau. La
pompe de dosage est immédiatement stoppée avec
signalisation sonore.
LED rouge «Too high» et «Too low» clignotent
simultanément = de l'eau a pénétré la platine
électronique du Controller, toutes les fonctions sont
inhibées : ouvrez le Controller (voir p. suivante),
séchez ou éventuellement nettoyez le circuit. Le
reset de l'appareil s'effectue par une coupure de
l'alimentation. Si les deux LED s'allument à nouveau
après quelques heures, des sels conducteurs sur la
platine ou de l'humidité résiduelle peuvent en être la
raison. L'appareil doit être retourné en réparation.