LEDs on Controller 507
Red LED „Too low" is flashing = The metering pump
ran for more than 10 minutes and was switched off.
The normal water level has not been reached. The
reason could be an empty storage tank, a kinked
or clogged hose, a defective pump or air in the
pump, a soiled, defective or disconnected operating
sensor (2). To reset the controller, briefly disconnect
the power supply.
Yellow LED „Pump on" is lit = The connected
metering pump (1) is operating. The LED is lit, and
the pump operates for a least 10 seconds, thus
preventing frequent operation.
Green LED „Level" is lit = The operating sensor (2)
reports water contact. If the operating sensor
does not detect water for 3 seconds (LED off), the
metering pump will start operating.
Controller 507 et diodes LED
LED rouge «Too low» clignote = la pompe de dosage
a dépassé un temps de fonctionnement de 10 min.
et se trouve stoppée. Le niveau d'eau normal n'a
pas été atteint. Raisons possibles : réservoir vide,
tuyau encrassé ou plié, pompe défectueuse, capteur
de régulation (2) décroché ou défectueux. Reset par
coupure de l'alimentation.
LED jaune «Pump on» allumée = pompe de
dosage (1) en fonctionnement durant un temps mini
de 10 sec. ce qui évite une trop grande fréquence
de mises en service.
LED verte «Level» allumée = signalisation de
présence d'eau sur le capteur de régulation (2).
Si le capteur ne signale plus d'eau durant plus
de 3 sec.(LED éteinte), la pompe de dosage est
actionnée.