Turning the potentiometer clockwise increase the
corresponding
readout
decreases the readout.
Repeat the procedure for the probe labeled OUT
(blue) using the corresponding potentiometer
labeled OUT for adjustment.
In case of gross deviations or erratic readings it is
advisable to first try the procedure again with fresh
batteries installed in the unit. The potentiometer (2)
is for gross range calibration and sets the range for
both electrodes this should only be used if the ppm
deviates by more than 100 ppm from the reference
solution on both probes. In such an instance the
gross range should be set to roughly 300ppm and
the calibration procedure outlined above should be
followed again.
and
counter-clockwise
Si le potentiomètre est ajusté dans le sens horaire,
la mesure en ppm augmente. S'il est ajusté dans le
sens anti-horaire, la mesure diminue.
Renouvelez la mesure avec le capteur "OUT" (bleu)
en utilisant si besoin le potentiomètre marqué
"OUT".
En cas de grandes différences de mesure
ou de mesures trop imprécises, renouvelez
l'étalonnage après le remplacement des batteries.
Le potentiomètre (2) est destiné à l'ajustage plus
grossier et concerne l'étalonnage simultané des
deux capteurs. Ce potentiomètre ne devrait être
ajusté que pour des valeurs de mesure ppm déviant
de plus de 100 ppm. Dans ce cas, il est nécessaire
d'indexer la valeur de mesure à environ 300 ppm
puis de reprendre toutes les étapes d'étalonnage
ci-dessus.