8.6.1 PRE-TARE function with 5 memories

Owing to this function it is possible to store 5 Pre-Tare values (e.g. for different
wheelchairs), and then call up them if necessary.
Saving PRE-Tare values:
To enable a later calling up values from the memory, they are to be entered into the
memory first. It is carried out in the following way:
The scale plate is unloaded, and the scale display is showing 0,0 kg.
Put a weight, whose value is to be saved (e.g. an empty wheelchair), on the scale
plate and wait until a stable weight display is shown.
Press the M1-5 key repeatedly until the display will show the „ni" (M) symbol.
Press the key with digit (1..5) shortly to indicate which number a value is to be
saved under. The previously displayed weight value is flashing for 3 seconds.
When the flashing is finished, press again the key with digit previously pressed and
the weighing value is saved in the memory (short audible signal).
By pressing the CLEAR key, the scale will return to the weighing mode without
saving the value.
The display will show the real value of the weight placed on the scale plate. When
the weight is removed, the display will show 0,0 kg.
Calling up the PRE-Tare value from memory:
Press the PRE-Tare key repeatedly until the display will show the „ni" (M) symbol.
Pressing the key with digit (1..5) will display the flashing weight value saved there.
The small arrow, additionally shown on the display, is pointing to the „PRE-TARE"
symbol. By pressing another key with digit (1..5), the appropriate also flashing
weight value will be displayed. Press the ENTER key and the value will be accepted
and shown on the display as the PRE-Tare value with minus sign.
Now you can place e.g. a person in a wheelchair or on the transportation stretcher on
the scale, and only a person's weight will be displayed.
To return to the normal weighing mode when the scale plate is unloaded, press the
PRE-Tare key shortly again. This will also result in disappearing the small arrow
pointing the „PRE-TARE" symbol.