Special Adjustments (For Service Personal Only); Adjusting The Level Sensor (Only 97009) - Loctite 97007 Bedienungsanleitung

2 l-halbautomatisches dosiersystem
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

8

Special Adjustments (for Service Personal only)

8.1
Adjusting the Level Sensor
Pay attention to the relevant safety regulations when working at
the open controller.
The level sensor must be adjusted
according to the type of product used and
the size of the bottle in order to function
properly.
• Loosen the (3) reservoir knobs 2 and
remove lid 1.
• Remove the screws and open the
housing.
• Empty a bottle of the product you use.
• Leave as much residue in the bottle as is required in order to prevent air getting into the
product feedline.
• Insert the product bottle. For small bottles use bottle nesting block 14.
• Check that the product bottle inserted into the bottle-nesting block is pressed against the level
sensor. Only then the correct adjustment of the level sensor is possible.
For product bottles, which do not fit into
the bottle-nesting block, rotate the level
sensor out or in so that it clamps the
product bottle.
If necessary, press the product bottle with
a suitable aid against the level sensor.
(only 97009)
23