Adjustment controls and seals

Possible seals:
enforced, and

6.9 Checking the balance verification settings

For the adjustment, the balance must be switched over to service mode.
In the service mode the parameters of the balance can be modified.
If this happened accidentally, please contact KERN.
In calibrated scales the service mode is locked individually for each switch. To
disable the access lock, destroy the seal and confirm the swatch. For position of
switch see chpt. 6.8.1.
Attention:
After destruction of the seal the balance must be re-verified by an authorised agency
and a new verification wire/seal mark fitted before it can be reused for applications
subject to verification.
18
or
1. Seal mark 1
2. Cover
3. Verification switch
4. Verification wire
RPB_D-BA-e-1726