4. Ensure that there are no objects on the weighing plate. Wait for stability display,
then press
. The required adjustment weight is shown.
5. Place the adjustment weight carefully in the centre of the weighing pan, press
and the adjusting process will be started.
6. The adjustment process has been successfully completed. Take away
adjustment weight.
7. Press
and the scales will return to weighing mode.
In case of an adjustment error or incorrect adjusting weight the display will
show an error message (SPAN / FAKL), repeat adjustment process.