8. Servicing/recalibration
Have the scale recalibrated by authorised persons in accordance with national legal re-
quirements. The year of initial calibration can be found behind the CE symbol on the rating
plate above the number of the appointed office, 0109 (Hessen Weights & Measures Of-
fice). Recalibration is essential if one or more security marks are damaged or the calibra-
tion counter does not match the number on the applicable calibration counter mark (cf.
page 17).
We recommend having your scale serviced by your local service agent before having it
recalibrated. seca Customer Service will be pleased to give you further assistance.

Changing the batteries

bAtt
When
or
6 type AA 1.5 V batteries.
– Carefully turn over the scale so that the
base of the scale is accessible.
– Push the battery compartment closure
towards "Battery" and open the cover.
– Take out the holder and batteries.
– Remove the used batteries and insert
new batteries in the holder.
Check that the polarity is correct.
– Place the holder and batteries in the
battery compartment.
– Close the cover.
– Then turn the scale back up the right
way.

9. Technical data

Maximum load
Graduations
0 up to 150 kg
150 kg up to 200 kg:
Dimensions scale
width:
height:
depth:
Dimensions weighing platform
width:
height:
depth:
Height of figures
Model 861
appear in the display, you should change the batteries. You need
200 kg
50 g
100 g
292 mm / 11.5 ins
244
409 mm / 16.1 ins
292 mm / 11.5 ins
75
300 mm / 11.8 ins
20 mm
+10
+0.4
mm / 9.6
ins
+10
+0.4
mm / 2.8
ins