APPENDICE D
Appendix D
The equipment is intended for use in the electromagnetic environment specified below. The customer or the end user should assure
Emission test
RF emission - CISPR 11
RF emission - CISPR 11
Harmonic emission IEC 61000-
3-2
Voltage fluctuation/flicker
emission IEC 61000-3-3
The equipment is intended for use in the electromagnetic environment specified below. The customer or the end user should assure
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical Fast
Transient/Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage Dips, Short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
IEC 60601-1-2 Table 1
Guidance and manufacturer's declaration - electromagnetic emissions
that it is used in such an environment
Compliance
Group 1
Therefore its RF emissions are very low and are not likely to cause any
Class B
The device is suitable for use in all establishments, including domestic
Class A
establishments and those directly connected to the public low-voltage power
Complies
IEC 60601-1-2 Table 2
Guidance and manufacturer's declaration - electromagnetic immunity
that it is used in such an environment
IEC 60601
Test level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for I/O lines
±1 kV line to line
±2 kV lines to earth
<5% Ut
(>95% dip in Ut ) for 0.5
cycle
40% Ut
(60% dip in Ut ) for 5 cycles
70% Ut
(30% dip in Ut ) for 5 cycles
5% Ut
(>95% dip in Ut )
for 5 s
3 A/m
Electromagnetic environment - guidance
The device uses RF energy only for its internal function.
interference in nearby electronic equipment
supply network that supplies buildings used for domestic purposes
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply lines
Not applicable
±1 kV line to line
±2 kV lines to earth
5% Ut
(>95% dip in Ut ) for 0.5 cycle
40% Ut
(60% dip in Ut ) for 5 cycles
70% Ut
(30% dip in Ut ) for 5 cycles
5% Ut
(>95% dip in Ut )
for 5 s
3 A/m
109
Electromagnetic environment -
guidance
Floor should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%
Mains power quality should be that of
a typical commercial or hospital
environment
Mains power quality should be that of
a typical commercial or hospital
environment
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the device
requires continued operation during
power mains interruptions, it is
recommended that the device be
powered from an Uninterruptible
Power Supply or Battery
Power frequency magnetic fields
should be at levels characteristics of
a typical location in a typical
commercial or hospital environment