Sterilization of Fiber-Optic Probe:
After each treatment, the fiber-optic probe should be disinfected or sterilized.
Wrap completely in an autoclave sterilization pouch, then autoclave at 132°C /75
psi for 4 minutes by using a pre-vacuum sterilizer to get maximum results. Allow
the fiber-optic probe to air-dry for 20-30 minutes in the sterilization pouch before
removing.
IMPORTANT :
Disconnect the fiber-optic probe from the handpiece and autoclave by itself.
IMPORTANT :
Every component of this device is reusable, but only the fiber-optic probe
needs to be disinfected or sterilized. Other components can not be
disinfected or sterilized.
Product Cleaning:
If there is any dirt or stains on the surface of control panel, handpiece or other
components, please use a cloth to wipe clean. If it is hard to clean, please use
a neutral cleansing agent. 70% isopropyl alcohol is recommended.
IMPORTANT :
Never use a solvent or volatile oil for cleaning; this can lead to fading
of the print on the casing or the metal ring falling off from the probe.