Trouble Shooting
Switch oft the device before any procedure and
unplug the power cable from the socket.
PROBLEM
CAUSE
The device does
The medication
not nebulise or
has not been
nebulises insuffi-
in-serted in the
ciently.
nebuliser.
The nebuliser
has not been
prop-erly fitted.
The nebuliser
nozzle is clog-
ged.
The air pipe is
not correctly
con-nected to
the device
The air pipe is
bent, damaged
or twisted
24
SOLUTION
Pour the right amount of me-
dication in the nebuliser.
Disassemble the nebuliser
and reassemble it correctly
as shown in the assembly
diagram on the cover.
Disassemble the nebuliser,
remove the nozzle and per-
form cleaning operations
Failure to remove medi-
cation deposits from the
nebuliser affects its efficien-
cy and operation Strictly
comply with the instructions
contained in the "Cleaning,
Sanitization and Disinfec-
tion" chapter.
Ensure the device's accesso-
ries are properly connected
to the air intake (see assem-
bly diagram on the cover).
Unwind the pipe and ensure
it is not crushed or punctu-
red. lf necessary replace it.